GB/T 29477-2012: General technical specification for mobile laboratory shelter
GB/T 29477-2012
GB
NATIONAL STANDARD OF THE
PEOPLE’S REPUBLIC OF CHINA
ICS 19.020
J 04
General technical specification for mobile laboratory shelter
ISSUED ON: DECEMBER 31, 2012
IMPLEMENTED ON: JULY 31, 2013
Issued by: General Administration of Quality Supervision, Inspection and
Quarantine of the People's Republic of China;
Standardization Administration of the People's Republic of China.
Table of Contents
Foreword ... 3
1 Scope ... 4
2 Normative references ... 4
3 Terms and definitions ... 5
4 Classification ... 5
5 Technical requirements ... 5
6 Test methods ... 12
7 Inspection rules ... 17
8 Packaging, marking, storage, transportation, technical documentation ... 19
General technical specification for mobile laboratory shelter
1 Scope
This Standard specifies the terms and definitions, classification, technical requirements,
test methods, inspection rules, packaging, marking, storage, transportation, technical
documents and other general technical requirements for mobile laboratory shelters.
This Standard applies to mobile laboratory shelters (hereinafter referred to as "shelters")
used on land.
2 Normative references
The following referenced documents are indispensable for the application of this
document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
GB 1589-2004, Limits of dimensions, axle load and masses for road vehicles
GB 7258-2012, Safety specifications for power-driven vehicles operating on roads
GB/T 13306, Plates
GB 14023-2011, Vehicles, boats and internal combustion engine -- Radio
disturbance characteristics -- Limits and methods of measurement for the protection
of off-board receivers
GB 16297-2004, General technical specification for mobile laboratory shelter
GB/T 18655-2010, Vehicles, boats and internal combustion engines -- Radio
disturbance characteristics -- Limits and methods of measurement for the protection
of on-board receivers
GB 19489, Laboratories -- General requirements for biosafety
GB/T 23334, Open-type roof escape hatch on bus
GB/T 25480-2010, Basic environmental conditions and testing methods for
transportation and storage of instruments
GB/T 29473-2012, Classification, code and marking of mobile laboratory
GB/T 29474-2012, General technical specification for the interior decoration
material of mobile laboratory
GB/T 29479-2012, General requirements of mobile laboratory
GJB 2093-1994, General test methods for military shelter
GJB 6109-2007, General specification for military shelter
JT/T 389-2010, Specification for van trailers
QC/T 452-1999, General technical conditions stay car
QC/T 484-1999, Automobile -- Paint Coating
YY 0569, Class II biological safety cabinets
3 Terms and definitions
For the purposes of this document, the terms and definitions defined in GB/T 29479-
2012 and GB/T 29473-2012apply.
4 Classification
4.1 According to different modes of movement, the shelters are divided into self-
propelled mobile shelters, trailer-type mobile shelters, and square shelter-type mobile
shelters.
4.2 According to different functions, the shelters are divided into physical mobile
shelters, chemical mobile shelters, biological mobile shelters, and comprehensive
mobile shelters.
5 Technical requirements
5.1 Environmental adaptability
5.1.1 The shelter shall be able to work normally under the following conditions:
a) Ambient temperature: -40℃ ~ +40℃. If necessary, the adaptable ambient
temperature of the shelter can be specified separately as needed. However, it shall
be clearly marked on the shelter sign;
b) Relative air humidity: below 95% (ambient temperature is 25℃);
c) Highways of Class IV and above.
5.1.2 The shelter shall have good impact resistance. It shall be able to work normally
after the impact test under the test conditions of peak acceleration of (100±20) m/s2 and
pulse duration of (16 ±2) ms.
5.6 Safety
5.6.1 The safety of self-propelled mobile shelters shall comply with the regulations of
GB 7258-2012.
5.6.2 The safety of the trailer-mounted mobile shelter shall comply with the regulations
of JT/T 389-2010.
5.6.3 The safety of the shelter-type mobile shelter shall comply with the regulations of
GJB 6109-2007.
5.6.4 The safety of the shelter shall also meet the following requirements:
a) Emergency exits shall be designed to facilitate emergency exit of personnel;
b) Smoke alarms and fire extinguishers shall be equipped, as well as toxic and
harmful gas alarms;
c) Leakage protection devices and grounding devices shall be equipped. When the
ambient temperature is 15℃ ~35℃ and the relative humidity is 45%~75%, the
cold insulation resistance value of each circuit of the electrical circuit to the
ground and between each other shall not be less than 2 MΩ.
5.6.5 The safety of biological mobile shelters shall also comply with the regulations of
GB 19489.
5.7 Design and structure
5.7.1 General requirements
The structure of the shelter shall meet the needs of different calibration and testing work.
According to functional needs, the shelter shall adopt the design structure of “three
areas, two barriers”. By designing two isolation barriers in the shelter, the auxiliary
equipment area, experimental operation area, and data processing area are relatively
independent to meet the development of various functional subsystems of the shelter.
The layout design in each area shall also comply with ergonomics principles.
5.7.2 Dimensions
5.7.2.1 The outer size limits of self-propelled and towed shelters shall be in accordance
with the provisions of GB 1589-2004. The outer dimensions of the square shelter are in
accordance with the provisions of GJB 6109-2007.
5.7.2.2 The dimensional deviation and geometric tolerance of the shelter shall meet the
following requirements.
a) Length: When the shelter length L≤4000 mm, the length limit deviation is ±5 mm;
When the shelter length L >4000 mm, the length deviation shall not exceed 0.15% of
the shelter length;
b) Width: the limit deviation is ±8 mm;
c) Height on the longitudinal center plane: the limit deviation is ±8 mm.
5.7.3 Doors, windows, openings
A reasonable number of doors, windows and openings shall be set up according to the
needs of the professional shelter. All shall ensure flexible opening and closing and
reliable operation. Limiting mechanisms or fixing devices shall be provided.
5.7.4 Door lock
All door locks must be flexible to open and close. When locked from outside the shelter,
the door can be opened from inside the shelter without tools.
5.7.5 Base plate load
The base plate bears the following static loads without plastic deformation or damage:
a) Uniform load: 1.5 kN/m2;
b) Concentrated load: when the area is 500 mm×500 mm, the load is 3 kN.
5.7.6 Roof plate load
The roof plate bears the following static loads without plastic deformation or damage:
a) Uniformly distributed load: 1 kN/m2;
b) Concentrated load: when the area is 300 mm×600 mm, the load is 1.5 kN.
5.7.7 Foot pedal
The shelter shall be equipped with auxiliary facilities for entering the shelter and getting
to the roof. There shall be no plastic deformation or damage when bearing a vertical
downward static load of 1.8 kN.
5.7.8 Leveling mechanism
Necessary supporting and balancing devices shall be provided. The accuracy can ensure <...
Get QUOTATION in 1-minute: Click GB/T 29477-2012
Historical versions: GB/T 29477-2012
Preview True-PDF (Reload/Scroll if blank)
GB/T 29477-2012: General technical specification for mobile laboratory shelter
GB/T 29477-2012
GB
NATIONAL STANDARD OF THE
PEOPLE’S REPUBLIC OF CHINA
ICS 19.020
J 04
General technical specification for mobile laboratory shelter
ISSUED ON: DECEMBER 31, 2012
IMPLEMENTED ON: JULY 31, 2013
Issued by: General Administration of Quality Supervision, Inspection and
Quarantine of the People's Republic of China;
Standardization Administration of the People's Republic of China.
Table of Contents
Foreword ... 3
1 Scope ... 4
2 Normative references ... 4
3 Terms and definitions ... 5
4 Classification ... 5
5 Technical requirements ... 5
6 Test methods ... 12
7 Inspection rules ... 17
8 Packaging, marking, storage, transportation, technical documentation ... 19
General technical specification for mobile laboratory shelter
1 Scope
This Standard specifies the terms and definitions, classification, technical requirements,
test methods, inspection rules, packaging, marking, storage, transportation, technical
documents and other general technical requirements for mobile laboratory shelters.
This Standard applies to mobile laboratory shelters (hereinafter referred to as "shelters")
used on land.
2 Normative references
The following referenced documents are indispensable for the application of this
document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
GB 1589-2004, Limits of dimensions, axle load and masses for road vehicles
GB 7258-2012, Safety specifications for power-driven vehicles operating on roads
GB/T 13306, Plates
GB 14023-2011, Vehicles, boats and internal combustion engine -- Radio
disturbance characteristics -- Limits and methods of measurement for the protection
of off-board receivers
GB 16297-2004, General technical specification for mobile laboratory shelter
GB/T 18655-2010, Vehicles, boats and internal combustion engines -- Radio
disturbance characteristics -- Limits and methods of measurement for the protection
of on-board receivers
GB 19489, Laboratories -- General requirements for biosafety
GB/T 23334, Open-type roof escape hatch on bus
GB/T 25480-2010, Basic environmental conditions and testing methods for
transportation and storage of instruments
GB/T 29473-2012, Classification, code and marking of mobile laboratory
GB/T 29474-2012, General technical specification for the interior decoration
material of mobile laboratory
GB/T 29479-2012, General requirements of mobile laboratory
GJB 2093-1994, General test methods for military shelter
GJB 6109-2007, General specification for military shelter
JT/T 389-2010, Specification for van trailers
QC/T 452-1999, General technical conditions stay car
QC/T 484-1999, Automobile -- Paint Coating
YY 0569, Class II biological safety cabinets
3 Terms and definitions
For the purposes of this document, the terms and definitions defined in GB/T 29479-
2012 and GB/T 29473-2012apply.
4 Classification
4.1 According to different modes of movement, the shelters are divided into self-
propelled mobile shelters, trailer-type mobile shelters, and square shelter-type mobile
shelters.
4.2 According to different functions, the shelters are divided into physical mobile
shelters, chemical mobile shelters, biological mobile shelters, and comprehensive
mobile shelters.
5 Technical requirements
5.1 Environmental adaptability
5.1.1 The shelter shall be able to work normally under the following conditions:
a) Ambient temperature: -40℃ ~ +40℃. If necessary, the adaptable ambient
temperature of the shelter can be specified separately as needed. However, it shall
be clearly marked on the shelter sign;
b) Relative air humidity: below 95% (ambient temperature is 25℃);
c) Highways of Class IV and above.
5.1.2 The shelter shall have good impact resistance. It shall be able to work normally
after the impact test under the test conditions of peak acceleration of (100±20) m/s2 and
pulse duration of (16 ±2) ms.
5.6 Safety
5.6.1 The safety of self-propelled mobile shelters shall comply with the regulations of
GB 7258-2012.
5.6.2 The safety of the trailer-mounted mobile shelter shall comply with the regulations
of JT/T 389-2010.
5.6.3 The safety of the shelter-type mobile shelter shall comply with the regulations of
GJB 6109-2007.
5.6.4 The safety of the shelter shall also meet the following requirements:
a) Emergency exits shall be designed to facilitate emergency exit of personnel;
b) Smoke alarms and fire extinguishers shall be equipped, as well as toxic and
harmful gas alarms;
c) Leakage protection devices and grounding devices shall be equipped. When the
ambient temperature is 15℃ ~35℃ and the relative humidity is 45%~75%, the
cold insulation resistance value of each circuit of the electrical circuit to the
ground and between each other shall not be less than 2 MΩ.
5.6.5 The safety of biological mobile shelters shall also comply with the regulations of
GB 19489.
5.7 Design and structure
5.7.1 General requirements
The structure of the shelter shall meet the needs of different calibration and testing work.
According to functional needs, the shelter shall adopt the design structure of “three
areas, two barriers”. By designing two isolation barriers in the shelter, the auxiliary
equipment area, experimental operation area, and data processing area are relatively
independent to meet the development of various functional subsystems of the shelter.
The layout design in each area shall also comply with ergonomics principles.
5.7.2 Dimensions
5.7.2.1 The outer size limits of self-propelled and towed shelters shall be in accordance
with the provisions of GB 1589-2004. The outer dimensions of the square shelter are in
accordance with the provisions of GJB 6109-2007.
5.7.2.2 The dimensional deviation and geometric tolerance of the shelter shall meet the
following requirements.
a) Length: When the shelter length L≤4000 mm, the length limit deviation is ±5 mm;
When the shelter length L >4000 mm, the length deviation shall not exceed 0.15% of
the shelter length;
b) Width: the limit deviation is ±8 mm;
c) Height on the longitudinal center plane: the limit deviation is ±8 mm.
5.7.3 Doors, windows, openings
A reasonable number of doors, windows and openings shall be set up according to the
needs of the professional shelter. All shall ensure flexible opening and closing and
reliable operation. Limiting mechanisms or fixing devices shall be provided.
5.7.4 Door lock
All door locks must be flexible to open and close. When locked from outside the shelter,
the door can be opened from inside the shelter without tools.
5.7.5 Base plate load
The base plate bears the following static loads without plastic deformation or damage:
a) Uniform load: 1.5 kN/m2;
b) Concentrated load: when the area is 500 mm×500 mm, the load is 3 kN.
5.7.6 Roof plate load
The roof plate bears the following static loads without plastic deformation or damage:
a) Uniformly distributed load: 1 kN/m2;
b) Concentrated load: when the area is 300 mm×600 mm, the load is 1.5 kN.
5.7.7 Foot pedal
The shelter shall be equipped with auxiliary facilities for entering the shelter and getting
to the roof. There shall be no plastic deformation or damage when bearing a vertical
downward static load of 1.8 kN.
5.7.8 Leveling mechanism
Necessary supporting and balancing devices shall be provided. The accuracy can ensure <...